MooneyCheck - Drive
Mooney Viscometer controlled by Personal Computer for the performance of Viscosity, Scorch and Stress Relaxation tests


Standards the instrument complies with:   
ASTM D1646;   ASTM D3346;   GOST 10722-76;   ISO 289-1;   ISO 289-2;   ISO 289-4;   JIS K_6300-1;    

 
Gibitre Instruments' Mooney viscometer measures Mooney viscosity, pre-polymerization (scorch), and stress relaxation in polymers and rubber compounds. The test determines the torque required to keep a rotor turning in a cylindrical chamber filled with rubber under controlled pressure, temperature, and speed—the resistance is the Mooney viscosity. It belongs to the suite of laboratory tools for the precise measurement of materials where repeatability is essential.

Why choose it: test chamber and rotor conform to ISO 289-1, ISO 289-2, ISO 289-4 and ASTM D1646, ASTM D3346; precise control via independent temperature controllers (0.1 °C resolution) and a compressed-air cooling circuit to minimize time between measurements; self-calibration with internal weight, pneumatic rotor ejection, touchscreen interface, and status lights for fast, safe operation.

For data management, Mooneycheck supports barcode sample ID and, with Datagest, full management of the Gibitre SQL database.

Options ext

Overview 
The MooneyCheck - Drive is a Mooney Viscometer, which permits to measure the Mooney viscosity, Pre-curing (scorch) and Stress Relaxation Characteristics of uncompounded or compounded rubbers In conformity with ISO 289-1, ISO 289-2, ISO 289-3, ASTM D 1646 Standards.
The test is carried out by measuring the torque, which has to be applied under specified temperature and pressure conditions, in order to rotate a metal disc at 2rpm speed in the cylindrical test chamber filled with rubber.
The resistance offered by the rubber to this rotation is expressed as the Mooney viscosity of the test piece.


Why choose Mooneycheck - Drive? 
• Entirely developed and produced in Italy
• Test chamber and rotor compliant with international standards
• Pneumatic Rotor Ejection
• Primary brand components
• Internal weight for self-calibration of the torque measurement
• Independent thermoregulators with a resolution of 0.1 °C
• Interactive touch-screen display for instrument control
• Optional control of rotor rotation speed
• Easily accessible test chamber with transparent safety panel and safety lock.
• Designed for eventual application of a fume aspiration system.
• Full license of MooneyCheck_10 software optimized for barcode sample identification
• Full license of the Datagest software for the complete management of the Gibitre SQL Standard Database




The new 'Drive' generation 
The new generation of 'Drive' instruments is the result of 40 years of experience in the measurement of rheology of rubber and incorporates the most modern technologies regarding mechanical construction, measurement sensors and control software.

MooneyCheck_10 software 
The MooneyCheck_10 program connects to the Gibitre SQL database to store test Results and Curves within the Datagest data management program.
The program allows you to:
• Quickly identify the sequence of tests to be performed
• Automatically adjust the instrument according to the required test conditions
• Produce the test Curves (Viscosity, Scorch, Stress Relaxation, Temperature)
• Compare the results with the tolerance limits
• Elaborate statistical analysis (Carta-X, Gaussiana, Media, Dev St., Max, Min, Cp, Cpk)
• Produce customized reports


Calculated Results 
The software allows you to prepare customized test procedures that include all the calculations required by the standards and product specifications. The dedicated page describes all the available calculations.



Option: Variable Rotor Rotation Speed 
Optional configuration of the MooneyCheck-Drive instrument for managing the rotor speed during the test.
• The configuration allows to define the Rotation Speed as a configurable parameter of the Test Procedure.
• The speed can be set between 0.01 and 20 RPM.
• The execution of tests at different speeds allow to evaluate the behavior of your product according to the shear rate and to simulate the behavior of the product in the conditions of use.


Instrument control Panel 
The instrument is equipped with a large touch-screen display with dimensions 10.2“.
The buttons on the display permit to start and stop the tests.
The display provides complete information about the status of the instrument: connection to the software, temperature of the dies, diagnostic of the sensors installed.


Rotor Pneumatic Ejection 
All Mooney Viscometers produced by Gibitre have a pneumatic ejection device for the rotor which simplifies the cleaning and replacement of the specimen between subsequent tests.



Light panel 
A light panel, installed in the front part of the instrument, changes the color and permits to check the status of the instrument from a distance.
The indicator light identifies the following statuses: Instrument ready, instrument under test, instrument setting test temperature.


Automatic Torque Calibration 
The Calibration weight integrated with the instrument permits the automatic Calibration of the torque sensor.
The Calibration is easily performed by pressing the Calibration button on the Control Display, which activates the pneumatic lifting of the weight.


Standard Calibration service for Mooney Viscometer 
The calibration is performed with reference to the requirements of ISO 289-1 standard.
The service includes:
• Ordinary maintenance of the instrument
• Replacement of the sealing of the lower die.
• Calibration of the temperature of upper and lower dies of instrument (at 100°C and 125°C).
• Calibration of the time for temperature recovery at test start
• Calibration of the torque reading (at 100 Mooney points)
• Calibration of the rotation speed of the rotor
• Final check with Standard Gibitre Compound
• Issue and e-mail shipment of the Calibration Certificate with traceability to primary standards.




Certified Rubber Samples For Mooney Test 
Pair of certified butyl rubber specimens (Isobutylene - Isoprene) for the verification of Mooney Viscosity.

The specimens are obtained, without rework, from a bale coded as IRM (Industry Reference Material) prepared by the working group ASTM D11.20 (task Group subcommittee on Compounding Materials and Procedures) in accordance with ASTM D 4678.


Datagest program : total Traceability 
The Datagest program is the Database Management Tool always installed in combination with all Gibitre instrument-control programs.
The program permits to:
• Select, filter, print, export and analyse the test results stored with all the instruments connected.
• Prepare test procedures by defining the test conditions and the results to be produced
• Set tolerance limits for each product by manual insertion or using the statistical analysis (mean and standard deviation) of saved results
• Prepare multi-instrument test reports


Gibitre Standard SQL Database 
All the Gibitre programs use a database with SQL structure for saving the results.
The database can be installed inside an SQL instance present on the company server or it can be installed on a PC connected to a measuring instrument. The installation of the Microsoft SQL service (Express version) is included in the delivery.




Industry 4.0 integration 
The instrument and the software have been specifically developed to optimize integration with other environments.
The database in SQL format and the Gibitre_Company_Connect program allows you synchronize your company management software with Gibitre database and to speed up the identification of the tests and to use bar-code readers or similar devices.


Preparation of Test Reports 
Version 10 programs make the following options available for preparing test reports:
• Selection of printing language
• Company logo insertion
• Print a single test / a Group of tests
• Print of Test Curves
• Print Numerical Results
• Print the Limits of Tolerance
• Print Statistical Results
• Print Legend with explanation of results
• Print the Operator Signature
• Print Customized Notes
• Preparation of PDF file

The program integrates a REPORT EDITOR that allows you to completely edit the report and prepare fully customized report formats.


Temperature regulation 
The temperature can be set between room temperature and 250°C.
The regulation of the temperature is performed using thermo-regulators with PID micro-processor and with 0.1°C resolution.
Independent temperature control units ensure sophisticate temperature control and easy replacement in case of failure.
Electrical heating resistances have been specifically designed for this instrument to ensure quick and efficient heating.




Safety devices 
The instrument is equipped with:
• Class 1 Safety switch, which prevents the closure of the dies if the safety panel is not closed.
• Safety Push-button
• Safety lock of the maintenance access door, which ensures safe usage even in non-standard operation conditions.
• CE Labelling


Easy cleaning of the instrument 
The instrument has been specifically developed for easy cleaning of the test chamber. Technical improvements are:
• Seal design to prevent the flowing of rubber below the rotor
• Easy rotor removing by pneumatic ejection
• The shaft housing hole is open from top to bottom to permit easy cleaning.
• The cleaning activity does not require regulation of rotor height


Development and production 
The instrument is totally developed and produced in the plant of Gibitre Instruments in Italy.
All the mechanical parts are produced in the company workshop using modern CNC machines.
Components and sensors from well-known brands are selected in order to ensure the maximum reliability in the measures
Internal trained personnel takes care of all the production stages: assembly, start-up, calibration, packing, shipment and installation.




 

Film Foils for Mooney Tests 
Polyamide or Polyester Film for the protection of the dies during the test.

Volumetric cutter 
Volumetric cutter for the preparation of test Samples with Constant Volume


Instrument Characteristics 
 
Type of tests performed 
Viscosity, Scorch, Stress Relaxation 
Torque sensor 
Capacity: 230 MU
Resolution: 0.01 MU
Linearity Error (%FS): +-0.25 
Temperature 
Between room temperature and +250 °C - Resolution 0.1 °C 
Frequency of rotation 
Standard Instrument: 2.00 r/min ± 0.02 r/min (0,209 rad/s ± 0,002 rad/s)
With variable speed option: adjustable between 0.01 to 20 r/min ± 1 % 
Software 
 
Numerical test data  
Viscosity test: MU_ini, t_ini, MU_min, t_min, Dt_MU, MU_4, MU_X, Dt(X-Y), S4, SX (X,Y=customer def.)
Scorch test: ts5, ts35, Dt_35-5, ts3, ts18, Dt_18-3, tsX, Dt_X-Y
Stress relaxation test: a, k, r, A, TX%  
Units 
Torque: Mooney Point (MU)
Time: minutes and seconds, minutes and minutes/100, seconds
Temperature: °C, °F 
Test curves 
Torque versus time curve, Log Mooney versus Log time of Stress Relaxation test, Upper and Lower Test Chamber temperatures 
Software 
 
Results storage 
The test result sand the curves are stored in the SQL Gibitre database which is installed in combination with the software 
Software usage Languages 
Italian, English, French, Spanish, German, Portuguese, Russian, Chinese, Japanese, Turkish, Polish, Czech 
Control Panel 
 
Data displayed 
Active connection to the software, motor on-off, temperatures of the dies, heating status, test running 


Type of device 
Capacitive-Touch-Screen display (permits the use with gloves) 
Characteristics 
Dimensions 10.2” 
Light Panel 
 
Permits to check from a distance the following statuses 
Instrument ready, instrument under test, instrument setting test temperature 
Safety Devices 
 
Safety Devices 
Class 1 Safety switch for main piston (Idem)
Safety Pushbutton
Safety lock of the maintenance access door
CE labelling 
Labelling 
CE Labelling 
Calibration 
 
Automatic Calibration system 
Calibration weight integrated with the instrument with pleumatic lifting system for automatic torque calibration 
Calibration Report 
Calibration report with traceability to primary standards in conformity with the Calibration requirements specified in ISO 289-1 standard 
Construction Characteristics 
 
Electronic Card 
Electronic card with STN 32F 429 micro-processor 
Power supply 
220 VAC ± 10%, 50-60 Hz ± 3, 4 A, single phase
110 VAC ± 10%, 60 Hz ± 3 on request 
Electrical power 
1100 Watt 
Compressed air 
6 bar 
Dimensions of instrument 
(Width x Depth x Height) 684 x 671 x 1419 mm 
Weight 
180 Kg 
Personal Computer (optional) 
 
Personal Computer (optional) 
Minimum Configuration: Intel Core i5 4 GB RAM.
Compatible Operating Systems: Windows 10 and 11;
Connection to the instrument via USB Cable (included) 
Options 
 
Volumetric Die Cutter 
For the preparation of samples with constant volume 
Film for testing 
 
Boxes with film foils 
Thickness: 0.023 mm
Dimension: 100 x 100 mm
Box Content: 500 foils
The boxex are supplied in couples. One box contains 500 standard foils. The other contains 500 foils with central hole for rotor insertion 


Gibitre Instruments
Via dell'industria, 18
Bergamo (Italy)
te. +39 035 461146
www.gibitre.it
info@gibitre.it

Copyright Gibitre Instruments
Mooney Viscometer controlled by Personal Computer for the performance of Viscosity, Scorch and Stress Relaxation tests
Campi di applicazione  >  Prove su pneumatici  >  Rheocheck MD - Drive
MooneyCheck - Drive
viscometer mooney drive: Mooney-Check粘度计驱动型,用于符合ISO 289、ASTM D1646和ASTM D3346标准的粘度测试。仪器全景视图。
Get a Quotation


MooneyCheck - Drive - 由个人电脑控制的门尼粘度计。
穆尼粘度计是一种旋转盘式粘度计,用于测量聚合物和橡胶混合物的穆尼粘度、焦化度及应力松弛特性。测试通过测量扭矩实现——在规定温度和压力条件下,需施加扭矩使金属盘在充满橡胶的圆柱形测试腔内以2rpm速度旋转。橡胶对旋转产生的阻力即为试样的穆尼粘度值。
主要特性
• 执行粘度、焦化及应力松弛测试
• 测试腔与转子符合国际标准
• 触摸屏控制操作界面
• 独立PID温控系统(分辨率0.1°C)
• 压缩空气冷却回路实现快速降温
• 内置校准砝码实现自动校准
• 气动转子弹出系统
• 远程状态监测指示灯
• 带安全传感器的透明防护罩
• 配备透明安全面板及安全锁的便捷测试腔
• 优化条码样品识别的Mooneycheck软件全功能授权

• Datagest软件完整授权,实现Gibitre SQL数据库全面管理
• 预装排烟装置
• 符合CE认证
配件
• 转子转速可在0.01至20 RPM间调节
• 小型转子
• 恒定体积样品切割器
• 测试期间保护模具用的聚酰胺或聚酯薄膜

Mooney Viscometer controlled by Personal Computer for the performance of Viscosity, Scorch and Stress Relaxation tests

Applications

Gibitre Instruments' Mooney viscometer measures Mooney viscosity, pre-polymerization (scorch), and stress relaxation in polymers and rubber compounds. The test determines the torque required to keep a rotor turning in a cylindrical chamber filled with rubber under controlled pressure, temperature, and speed—the resistance is the Mooney viscosity. It belongs to the suite of laboratory tools for the precise measurement of materials where repeatability is essential.

Why choose it: test chamber and rotor conform to ISO 289-1, ISO 289-2, ISO 289-4 and ASTM D1646, ASTM D3346; precise control via independent temperature controllers (0.1 °C resolution) and a compressed-air cooling circuit to minimize time between measurements; self-calibration with internal weight, pneumatic rotor ejection, touchscreen interface, and status lights for fast, safe operation.

For data management, Mooneycheck supports barcode sample ID and, with Datagest, full management of the Gibitre SQL database.

Options ext
Why Choose MooneyCheck - Drive?

Overview 

The MooneyCheck - Drive is a Mooney Viscometer, which permits to measure the Mooney viscosity, Pre-curing (scorch) and Stress Relaxation Characteristics of uncompounded or compounded rubbers In conformity with ISO 289-1, ISO 289-2, ISO 289-3, ASTM D 1646 Standards.
The test is carried out by measuring the torque, which has to be applied under specified temperature and pressure conditions, in order to rotate a metal disc at 2rpm speed in the cylindrical test chamber filled with rubber.
The resistance offered by the rubber to this rotation is expressed as the Mooney viscosity of the test piece.

viscometer mooney drive: 门尼粘度计。转子详图,符合 ISO ISO 289、ASTM D1646 和 ASTM D3346 标准。 viscometer mooney drive: 门尼粘度计。转子详图,符合 ISO ISO 289、ASTM D1646 和 ASTM D3346 标准。 viscometer mooney drive: 用热像仪检查试验箱温度

Why choose Mooneycheck - Drive? 

• Entirely developed and produced in Italy
• Test chamber and rotor compliant with international standards
• Pneumatic Rotor Ejection
• Primary brand components
• Internal weight for self-calibration of the torque measurement
• Independent thermoregulators with a resolution of 0.1 °C
• Interactive touch-screen display for instrument control
• Optional control of rotor rotation speed
• Easily accessible test chamber with transparent safety panel and safety lock.
• Designed for eventual application of a fume aspiration system.
• Full license of MooneyCheck_10 software optimized for barcode sample identification
• Full license of the Datagest software for the complete management of the Gibitre SQL Standard Database

viscometer mooney drive: 意大利精心制造,经久耐用
意大利精心制造,经久耐用

The new 'Drive' generation 

The new generation of 'Drive' instruments is the result of 40 years of experience in the measurement of rheology of rubber and incorporates the most modern technologies regarding mechanical construction, measurement sensors and control software.

viscometer mooney drive: 驱动版仪器型号的徽标。
驱动版仪器型号的徽标。
viscometer mooney drive: Mooney-Check 软件第10版。软件主界面显示叠加的测试曲线、数值结果及统计数据。
 
Software

MooneyCheck_10 software 

The MooneyCheck_10 program connects to the Gibitre SQL database to store test Results and Curves within the Datagest data management program.
The program allows you to:
• Quickly identify the sequence of tests to be performed
• Automatically adjust the instrument according to the required test conditions
• Produce the test Curves (Viscosity, Scorch, Stress Relaxation, Temperature)
• Compare the results with the tolerance limits
• Elaborate statistical analysis (Carta-X, Gaussiana, Media, Dev St., Max, Min, Cp, Cpk)
• Produce customized reports

viscometer mooney drive: Mooney-Check 软件第10版。软件主界面显示叠加的测试曲线、数值结果及统计数据。
Mooney-Check 软件第10版。软件主界面显示叠加的测试曲线、数值结果及统计数据。

Calculated Results 

The software allows you to prepare customized test procedures that include all the calculations required by the standards and product specifications. The dedicated page describes all the available calculations.

viscometer mooney drive: Gibitre 软件。结果计算公式。Mooney-Check 软件。Delta-Mooney 计算:设定时间 t1 和 t2 之间粘度值的变化。 viscometer mooney drive: Gibitre 软件。结果计算公式。Mooney-Check 软件。Delta-Mooney 计算:设定时间 t1 和 t2 之间粘度值的变化。 viscometer mooney drive: Gibitre 软件。计算结果的公式。穆尼检查松弛试验回归线计算公式。用双对数图表示。 viscometer mooney drive: Gibitre 软件。结果计算公式。Mooney-Check 软件灼烧试验计算 Delta_ts:粘度在 5 至 35 MU 之间上升所需的时间。

Option: Variable Rotor Rotation Speed 

Optional configuration of the MooneyCheck-Drive instrument for managing the rotor speed during the test.
• The configuration allows to define the Rotation Speed as a configurable parameter of the Test Procedure.
The speed can be set between 0.01 and 20 RPM.
• The execution of tests at different speeds allow to evaluate the behavior of your product according to the shear rate and to simulate the behavior of the product in the conditions of use.

viscometer mooney drive: Gibitre 软件。用于检查穆尼粘度计的 Mooney-Check 程序。不同转子速度下的测试运行列表。 viscometer mooney drive: Gibitre 软件。用于检查穆尼粘度计的 Mooney-Check 程序。不同转子速度下的测试运行列表。 viscometer mooney drive: Gibitre 软件。可定制的测试报告。报告可以用不同颜色标识测试,以识别相应的测试曲线。 viscometer mooney drive: 穆尼粘度计:转子速度可调。
viscometer mooney drive: 触摸屏控制面板
Developed for the user

Instrument control Panel 

The instrument is equipped with a large touch-screen display with dimensions 10.2“.
The buttons on the display permit to start and stop the tests.
The display provides complete information about the status of the instrument: connection to the software, temperature of the dies, diagnostic of the sensors installed.

viscometer mooney drive: 触摸显示屏:一切尽在指尖 viscometer mooney drive: 触摸显示屏:一切尽在指尖 viscometer mooney drive: Mooney-Check-Drive: 显示触摸屏,用于控制乐器。

Rotor Pneumatic Ejection 

All Mooney Viscometers produced by Gibitre have a pneumatic ejection device for the rotor which simplifies the cleaning and replacement of the specimen between subsequent tests.

viscometer mooney drive: 气动转子弹射细节 viscometer mooney drive: 气动转子弹射细节 viscometer mooney drive: 气动转子弹射细节 viscometer mooney drive: 气动转子弹射细节

Light panel 

A light panel, installed in the front part of the instrument, changes the color and permits to check the status of the instrument from a distance.
The indicator light identifies the following statuses: Instrument ready, instrument under test, instrument setting test temperature.

viscometer mooney drive: 远距离检查仪器状态 viscometer mooney drive: 远距离检查仪器状态 viscometer mooney drive: 实验室压力机:运行状态指示灯面板正视图。
viscometer mooney drive: 气动转子弹射顺序
Calibration

Automatic Torque Calibration 

The Calibration weight integrated with the instrument permits the automatic Calibration of the torque sensor.
The Calibration is easily performed by pressing the Calibration button on the Control Display, which activates the pneumatic lifting of the weight.

viscometer mooney drive: 自动扭矩校准
自动扭矩校准

Standard Calibration service for Mooney Viscometer 

The calibration is performed with reference to the requirements of ISO 289-1 standard.
The service includes:
• Ordinary maintenance of the instrument
• Replacement of the sealing of the lower die.
• Calibration of the temperature of upper and lower dies of instrument (at 100°C and 125°C).
• Calibration of the time for temperature recovery at test start
• Calibration of the torque reading (at 100 Mooney points)
• Calibration of the rotation speed of the rotor
• Final check with Standard Gibitre Compound
• Issue and e-mail shipment of the Calibration Certificate with traceability to primary standards.

Sample Calibration Report
viscometer mooney drive: 吉比特校准:完全符合标准
吉比特校准:完全符合标准

Certified Rubber Samples For Mooney Test 

Pair of certified butyl rubber specimens (Isobutylene - Isoprene) for the verification of Mooney Viscosity.

The specimens are obtained, without rework, from a bale coded as IRM (Industry Reference Material) prepared by the working group ASTM D11.20 (task Group subcommittee on Compounding Materials and Procedures) in accordance with ASTM D 4678.

viscometer mooney drive: 经过认证的测试样本,用于穆尼粘度测试。包装内含一对真空测试样本和识别标签。 viscometer mooney drive: 经过认证的测试样本,用于穆尼粘度测试。包装内含一对真空测试样本和识别标签。 viscometer mooney drive: 用于穆尼粘度计校准验证的认证样品。由 ASTM D11.20 工作组复合材料分会编制的被编码为 IRM(工业参考材料)的料包的校准证书和由 ASTM D11.20 工作组复合材料分会编制的被编码为 IRM(工业参考材料)的料包的校准证书。
viscometer mooney drive: Datagest软件第10版用于管理存档结果。结果选择页面的详细信息,附带测试报告预览。
Data Management

Datagest program : total Traceability 

The Datagest program is the Database Management Tool always installed in combination with all Gibitre instrument-control programs.
The program permits to:
Select, filter, print, export and analyse the test results stored with all the instruments connected.
Prepare test procedures by defining the test conditions and the results to be produced
Set tolerance limits for each product by manual insertion or using the statistical analysis (mean and standard deviation) of saved results
Prepare multi-instrument test reports

viscometer mooney drive: 对所有结果进行统一管理
对所有结果进行统一管理

Gibitre Standard SQL Database 

All the Gibitre programs use a database with SQL structure for saving the results.
The database can be installed inside an SQL instance present on the company server or it can be installed on a PC connected to a measuring instrument. The installation of the Microsoft SQL service (Express version) is included in the delivery.

viscometer mooney drive: SQL:全面保护您的数据安全
SQL:全面保护您的数据安全

Industry 4.0 integration 

The instrument and the software have been specifically developed to optimize integration with other environments.
The database in SQL format and the Gibitre_Company_Connect program allows you synchronize your company management software with Gibitre database and to speed up the identification of the tests and to use bar-code readers or similar devices.

viscometer mooney drive: 管理软件与实验室软件的整合
管理软件与实验室软件的整合

Preparation of Test Reports 

Version 10 programs make the following options available for preparing test reports:
• Selection of printing language
• Company logo insertion
• Print a single test / a Group of tests
• Print of Test Curves
• Print Numerical Results
• Print the Limits of Tolerance
• Print Statistical Results
• Print Legend with explanation of results
• Print the Operator Signature
• Print Customized Notes
• Preparation of PDF file

The program integrates a REPORT EDITOR that allows you to completely edit the report and prepare fully customized report formats.

viscometer mooney drive: 测试报告:您的名片 viscometer mooney drive: 测试报告:您的名片 viscometer mooney drive: Gibitre 程序。仪器软件可根据测试结果定制测试报告。用于选择报告模板和打印语言的页面。 viscometer mooney drive: 准备定制的报告格式

Temperature regulation 

The temperature can be set between room temperature and 250°C.
The regulation of the temperature is performed using thermo-regulators with PID micro-processor and with 0.1°C resolution.
Independent temperature control units ensure sophisticate temperature control and easy replacement in case of failure.
Electrical heating resistances have been specifically designed for this instrument to ensure quick and efficient heating.

viscometer mooney drive: Mooney-Check 粘度计软件。测试运行期间记录腔室温度的细节。
Mooney-Check 粘度计软件。测试运行期间记录腔室温度的细节。

Safety devices 

The instrument is equipped with:
Class 1 Safety switch, which prevents the closure of the dies if the safety panel is not closed.
• Safety Push-button
Safety lock of the maintenance access door, which ensures safe usage even in non-standard operation conditions.
CE Labelling

viscometer mooney drive: 安全第一
安全第一

Easy cleaning of the instrument 

The instrument has been specifically developed for easy cleaning of the test chamber. Technical improvements are:
• Seal design to prevent the flowing of rubber below the rotor
• Easy rotor removing by pneumatic ejection
• The shaft housing hole is open from top to bottom to permit easy cleaning.
• The cleaning activity does not require regulation of rotor height

viscometer mooney drive: 用于清洁的腔室快速拆卸顺序 - 第 1 阶段 viscometer mooney drive: 用于清洁的腔室快速拆卸顺序 - 第 1 阶段 viscometer mooney drive: 用于清洁的快速炉室拆卸程序 - 第 2 阶段 viscometer mooney drive: 用于清洁的快速炉室拆卸顺序 - 第 3 步

Development and production 

The instrument is totally developed and produced in the plant of Gibitre Instruments in Italy.
All the mechanical parts are produced in the company workshop using modern CNC machines.
Components and sensors from well-known brands are selected in order to ensure the maximum reliability in the measures
Internal trained personnel takes care of all the production stages: assembly, start-up, calibration, packing, shipment and installation.

viscometer mooney drive: 引以为豪的意大利制造
引以为豪的意大利制造

 

Accessories

Film Foils for Mooney Tests 

Polyamide or Polyester Film for the protection of the dies during the test.

viscometer mooney drive: 恒定体积样品 = 更高的重复性 viscometer mooney drive: 恒定体积样品 = 更高的重复性 viscometer mooney drive: 门尼粘度试验。试验准备。安装在转子上方和下方的两个试样及其腔室保护板。

Volumetric cutter 

Volumetric cutter for the preparation of test Samples with Constant Volume

viscometer mooney drive: 恒定体积样品 = 更高的重复性
恒定体积样品 = 更高的重复性

Technical specification

Reference Standards
Standards the instrument complies with:  ISO 289-1; ISO 289-2; ISO 289-4; ASTM D1646; ASTM D3346; GOST 10722-76; JIS K_6300-1;  

Instrument Characteristics

Type of tests performed  Viscosity, Scorch, Stress Relaxation  Torque sensor  Capacity: 230 MU
Resolution: 0.01 MU
Linearity Error (%FS): +-0.25 
Temperature  Between room temperature and +250 °C - Resolution 0.1 °C  Frequency of rotation  Standard Instrument: 2.00 r/min ± 0.02 r/min (0,209 rad/s ± 0,002 rad/s)
With variable speed option: adjustable between 0.01 to 20 r/min ± 1 % 

Software

Numerical test data   Viscosity test: MU_ini, t_ini, MU_min, t_min, Dt_MU, MU_4, MU_X, Dt(X-Y), S4, SX (X,Y=customer def.)
Scorch test: ts5, ts35, Dt_35-5, ts3, ts18, Dt_18-3, tsX, Dt_X-Y
Stress relaxation test: a, k, r, A, TX%  
Units  Torque: Mooney Point (MU)
Time: minutes and seconds, minutes and minutes/100, seconds
Temperature: °C, °F 
Test curves  Torque versus time curve, Log Mooney versus Log time of Stress Relaxation test, Upper and Lower Test Chamber temperatures 

Software

Results storage  The test result sand the curves are stored in the SQL Gibitre database which is installed in combination with the software  Software usage Languages  Italian, English, French, Spanish, German, Portuguese, Russian, Chinese, Japanese, Turkish, Polish, Czech 

Control Panel

Data displayed  Active connection to the software, motor on-off, temperatures of the dies, heating status, test running  Type of device  Capacitive-Touch-Screen display (permits the use with gloves)  Characteristics  Dimensions 10.2” 

Light Panel

Permits to check from a distance the following statuses  Instrument ready, instrument under test, instrument setting test temperature 

Safety Devices

Safety Devices  Class 1 Safety switch for main piston (Idem)
Safety Pushbutton
Safety lock of the maintenance access door
CE labelling 
Labelling  CE Labelling 

Calibration

Automatic Calibration system  Calibration weight integrated with the instrument with pleumatic lifting system for automatic torque calibration  Calibration Report  Calibration report with traceability to primary standards in conformity with the Calibration requirements specified in ISO 289-1 standard 

Construction Characteristics

Electronic Card  Electronic card with STN 32F 429 micro-processor  Power supply  220 VAC ± 10%, 50-60 Hz ± 3, 4 A, single phase
110 VAC ± 10%, 60 Hz ± 3 on request 
Electrical power  1100 Watt  Compressed air  6 bar  Dimensions of instrument  (Width x Depth x Height) 684 x 671 x 1419 mm  Weight  180 Kg 

Personal Computer (optional)

Personal Computer (optional)  Minimum Configuration: Intel Core i5 4 GB RAM.
Compatible Operating Systems: Windows 10 and 11;
Connection to the instrument via USB Cable (included) 

Options

Volumetric Die Cutter  For the preparation of samples with constant volume 

Film for testing

Boxes with film foils  Thickness: 0.023 mm
Dimension: 100 x 100 mm
Box Content: 500 foils
The boxex are supplied in couples. One box contains 500 standard foils. The other contains 500 foils with central hole for rotor insertion 
References
Gibitre Mooney Check
is currently installed in
35 countries
Request the List of top References
Download
Extended product brochure   
Brief product leaflet
General catalogue mail request

Calibration certificate
example
CE certificate
example
RheoCheck MD - Drive
Vedi anche: Reometro Rheocheck MD
See Also   >
Constant Volume Sample Cutter - Mooney
Vedi anche: 体积式模切机驱动装置:用于粘度计的穆尼体积式模切机全景。
See Also   >
Certified Samples for Mooney Test
Vedi anche: 经过认证的测试样本,用于穆尼粘度测试。包装内含一对真空测试样本和识别标签。
See Also   >
News
Call to action
Ask for quotation
a guided way to ask
for a quotation
Mail
info@gibitre.it
The good old e-mail
ready for every season
Phone
+39 035 460146
We welcome your calls Monday through Friday from 8:30 a.m. to 5:30 p.m. in your local time zone.
Newsletter
Subscribe
Stay updated
with the last news
of your sector
Local agent
from: -
See the whole list

In your country
in your language
at your service
Gibitre Instruments S.r.l., Via dell'industria 73, 24126 Bergamo Italia, pec@gibitre-instruments.it P. IVA: IT02827900164 CCIAA REA Bergamo: 326486 | ©2018 - 2026 Gibitre Instruments
Optimist.it
Welcome to the Gibitre Instruments web site
Please chose the preferred Cookie Management.
Benvenuti nel sito Gibitre Instruments
Per continuare scegliete la Gestione dei Cookie che preferite.
Cookies policy
For detailed informations about the use of cookies, please read our
Cookies Management Practice.

Gestione dei Cookies
Per maggiori informazioni sui cookie utilizzati consultate la nostra
Procedura sull'Uso dei Cookies.