MooneyCheck: Test Results
Below you can find the description of the test results produced by MooneyCheck software



 

MU_xx: Mooney Viscosity at set time 
The calculation returns the Mooney Viscosity after the set test time. Standard test time includes 1 minute pre-heating time and 4 minutes test time.

REFERENCE STANDARD: ISO 289-1 / ASTM D 1646

UNIT: MU (Mooney units)

CALCULATION PARAMETERS:
Test Time: t

EXPRESSION:
ML (1+t) - Example ML (1+4)


MU_min: Minimum Mooney Viscosity 
The calculation returns the minimum value of Mooney Viscosity over the complete test time.

UNIT: MU (Mooney units)

EXPRESSION:
MU_min




MU_ini: Initial Mooney Viscosity 
The calculation returns the Maximum Mooney viscosity at the beginning of test time.

UNIT: MU (Mooney units)

EXPRESSION:
MU_ini


S_xx: Viscosity reduction Rate at set time 
The calculation returns the Slope of curve of the Viscosity vs Time at the set test time.

UNIT: MU (Mooney units)

PARAMETERS:
Test Time: t

EXPRESSION:
S (1+t) - Example S(1+12)


Delta_MU_T1_T2: Variation of Mooney Viscosity  
Reference standard ISO 289-3
The calculation returns the difference between the Mooney viscosities recorded at running time t1 and t2.

UNIT: MU (mooney units)

PARAMETERS:
Test Time t1
Test Time t2

EXPRESSION:
Dt_MU_(t2-t1)




ts: Scorch Time 
OVERVIEW:
The pre-vulcanization characteristics provide a means of estimating how long compounded rubber can be maintained at high temperatures and remain processable.

REFERENCE STANDARD:
ISO 289-3 / ASTM D 1646

The calculation returns the time in minutes including warm-up time, for the viscosity to increase by a specified amount from the minimum value.

UNIT: mm:cc (minutes, minutes/100) or mm:ss (minutes, seconds)

PARAMETERS:
Torque Increase: MU

EXPRESSION:
ts_MU (example ts_35)


Dt_ts: Scorch Index 
The calculation returns the time needed for the viscosity to increase from the first to the second set values.

REFERENCE STANDARD: ASTM D 1646

UNIT: mm:cc (minutes, minutes/100) or mm:ss (minutes, seconds)

PARAMETERS:
Start Torque Value: ts1
Final Torque Value: ts2

EXPRESSION:
Dt_ts_(ts2-ts1) - Example: Dt_ts(35-5)


a (MSR): Mooney Stress-Relaxation Rate 
OVERVIEW:
Stress Relaxation test consists of determining the decay of the Mooney torque immediately after the determination of the Mooney viscosity. After abruptly stopping the rotor at the end of the Mooney viscosity measurement, the decrease in torque is recorded as a function of time. The rate of change of the torque is evaluated over a short time interval.

REFERENCE STANDARD: ISO 289-4

CALCULATION:
The calculation returns the slope of the linear regression line of the log(torque) versus log(time) plot over a specified time interval after stopping the rotor at the end of a Mooney viscosity measurement. According to ISO Standard the regression is calculated between 1.6 and 5 seconds from rotor stop.

UNIT: Mu/sec

PARAMETERS:
Time for Start of Regression Calculation: t1
Time for Stop of Regression Calculation: t2

EXPRESSION:
a(MSR) t1_t2 - Example: a(MSR) 1.6_5.0




A: Mooney Stress-Relaxation Area 
The calculation returns the area under the stress relaxation curve over a specified time interval after stopping the rotor at the end of a Mooney viscosity measurement.

REFERENCE STANDARD: ASTM D 1646

UNIT: Mu*sec

PARAMETERS:
Time for Start of the Calculation: t1
Time for Stop of the Calculation: t2

EXPRESSION:
A (t1_t2) - Example: A (1.6_5.0)



Gibitre Instruments
Via dell'industria, 18
Bergamo (Italy)
te. +39 035 461146
www.gibitre.it
info@gibitre.it

Copyright Gibitre Instruments
Below you can find the description of the test results produced by MooneyCheck software
Campi di applicazione  >  Prove su pneumatici  >  Rheocheck MD - Drive
MooneyCheck: Test Results
viscometer mooney formulas: Mooney viscometer: software calculation formulas: scorch time.

Below you can find the description of the test results produced by MooneyCheck software

Applications

MU_xx: Mooney Viscosity at set time 

The calculation returns the Mooney Viscosity after the set test time. Standard test time includes 1 minute pre-heating time and 4 minutes test time.

REFERENCE STANDARD: ISO 289-1 / ASTM D 1646

UNIT: MU (Mooney units)

CALCULATION PARAMETERS:
Test Time: t

EXPRESSION:
ML (1+t) - Example ML (1+4)

viscometer mooney formulas: Gibitre software. Formulas for calculating results. Mooney-Check software. Viscosity test. Mooney viscosity after set time.
Gibitre software. Formulas for calculating results. Mooney-Check software. Viscosity test. Mooney viscosity after set time.

MU_min: Minimum Mooney Viscosity 

The calculation returns the minimum value of Mooney Viscosity over the complete test time.

UNIT: MU (Mooney units)

EXPRESSION:
MU_min

viscometer mooney formulas: Gibitre software. Formulas for calculating results. Mooney-Check software. Viscosity test. Measurement of Mu_min: minimum viscosity value recorded during the test.
Gibitre software. Formulas for calculating results. Mooney-Check software. Viscosity test. Measurement of Mu_min: minimum viscosity value recorded during the test.

MU_ini: Initial Mooney Viscosity 

The calculation returns the Maximum Mooney viscosity at the beginning of test time.

UNIT: MU (Mooney units)

EXPRESSION:
MU_ini

viscometer mooney formulas: Gibitre software. Formulas for calculating results. Mooney-Check software. Viscosity test. Maximum viscosity upon activation of rotor rotation.
Gibitre software. Formulas for calculating results. Mooney-Check software. Viscosity test. Maximum viscosity upon activation of rotor rotation.
 

S_xx: Viscosity reduction Rate at set time 

The calculation returns the Slope of curve of the Viscosity vs Time at the set test time.

UNIT: MU (Mooney units)

PARAMETERS:
Test Time: t

EXPRESSION:
S (1+t) - Example S(1+12)

viscometer mooney formulas: Gibitre software. Formulas for calculating results. Mooney-Check software. Viscosity Test. Calculation of the slope of the viscosity curve at the set time.
Gibitre software. Formulas for calculating results. Mooney-Check software. Viscosity Test. Calculation of the slope of the viscosity curve at the set time.

Delta_MU_T1_T2: Variation of Mooney Viscosity  

Reference standard ISO 289-3
The calculation returns the difference between the Mooney viscosities recorded at running time t1 and t2.

UNIT: MU (mooney units)

PARAMETERS:
Test Time t1
Test Time t2

EXPRESSION:
Dt_MU_(t2-t1)

viscometer mooney formulas: Gibitre software. Formulas for calculating results. Mooney-Check software. Delta-Mooney calculation: Change in viscosity value between set times t1 and t2.
Gibitre software. Formulas for calculating results. Mooney-Check software. Delta-Mooney calculation: Change in viscosity value between set times t1 and t2.

ts: Scorch Time 

OVERVIEW:
The pre-vulcanization characteristics provide a means of estimating how long compounded rubber can be maintained at high temperatures and remain processable.

REFERENCE STANDARD:
ISO 289-3 / ASTM D 1646

The calculation returns the time in minutes including warm-up time, for the viscosity to increase by a specified amount from the minimum value.

UNIT: mm:cc (minutes, minutes/100) or mm:ss (minutes, seconds)

PARAMETERS:
Torque Increase: MU

EXPRESSION:
ts_MU (example ts_35)

viscometer mooney formulas: Gibitre software. Formulas for calculating results. Mooney-Check software. Scorch test (scorch). Calculation of ts: time needed to reach the set Viscosity value (e.g. ts_5).
Gibitre software. Formulas for calculating results. Mooney-Check software. Scorch test (scorch). Calculation of ts: time needed to reach the set Viscosity value (e.g. ts_5).

Dt_ts: Scorch Index 

The calculation returns the time needed for the viscosity to increase from the first to the second set values.

REFERENCE STANDARD: ASTM D 1646

UNIT: mm:cc (minutes, minutes/100) or mm:ss (minutes, seconds)

PARAMETERS:
Start Torque Value: ts1
Final Torque Value: ts2

EXPRESSION:
Dt_ts_(ts2-ts1) - Example: Dt_ts(35-5)

viscometer mooney formulas: Gibitre software. Formulas for calculating results. Mooney-Check software. Scorch test (scorch). Calculation of Delta_ts: time required for viscosity increase between 5 and 35 MU.
Gibitre software. Formulas for calculating results. Mooney-Check software. Scorch test (scorch). Calculation of Delta_ts: time required for viscosity increase between 5 and 35 MU.

a (MSR): Mooney Stress-Relaxation Rate 

OVERVIEW:
Stress Relaxation test consists of determining the decay of the Mooney torque immediately after the determination of the Mooney viscosity. After abruptly stopping the rotor at the end of the Mooney viscosity measurement, the decrease in torque is recorded as a function of time. The rate of change of the torque is evaluated over a short time interval.

REFERENCE STANDARD: ISO 289-4

CALCULATION:
The calculation returns the slope of the linear regression line of the log(torque) versus log(time) plot over a specified time interval after stopping the rotor at the end of a Mooney viscosity measurement. According to ISO Standard the regression is calculated between 1.6 and 5 seconds from rotor stop.

UNIT: Mu/sec

PARAMETERS:
Time for Start of Regression Calculation: t1
Time for Stop of Regression Calculation: t2

EXPRESSION:
a(MSR) t1_t2 - Example: a(MSR) 1.6_5.0

viscometer mooney formulas: Gibitre software. Formulas for calculating results. Mooney-Check software. Definition of times t1 and t2 for calculating the regression line of the Relaxation stress treatment. viscometer mooney formulas: Gibitre software. Formulas for calculating results. Mooney-Check software. Definition of times t1 and t2 for calculating the regression line of the Relaxation stress treatment. viscometer mooney formulas: Gibitre software. Formulas for calculating results. Mooney-Check. Formula for calculation of regression line of relaxation test. Representation in bilogarithmic diagram.

A: Mooney Stress-Relaxation Area 

The calculation returns the area under the stress relaxation curve over a specified time interval after stopping the rotor at the end of a Mooney viscosity measurement.

REFERENCE STANDARD: ASTM D 1646

UNIT: Mu*sec

PARAMETERS:
Time for Start of the Calculation: t1
Time for Stop of the Calculation: t2

EXPRESSION:
A (t1_t2) - Example: A (1.6_5.0)

viscometer mooney formulas: Gibitre software. Formulas for calculating results. Mooney-Check software. Calculation of energy in the Stress-Relaxation test.
Gibitre software. Formulas for calculating results. Mooney-Check software. Calculation of energy in the Stress-Relaxation test.
Download
General catalogue mail request

RheoCheck: Test Results
Vedi anche: Software Rheometro: Software calculation formulas. Delta tnagent.
See Also   >
News
Call to action
Mail
info@gibitre.it
The good old e-mail
ready for every season
Phone
+39 035 460146
We welcome your calls Monday through Friday from 8:30 a.m. to 5:30 p.m. in your local time zone.
Newsletter
Subscribe
Stay updated
with the last news
of your sector
Local agent
from: -
See the whole list

In your country
in your language
at your service
Gibitre Instruments S.r.l., Via dell'industria 73, 24126 Bergamo Italia, pec@gibitre-instruments.it P. IVA: IT02827900164 CCIAA REA Bergamo: 326486 | ©2018 - 2026 Gibitre Instruments
Optimist.it
Welcome to the Gibitre Instruments web site
Please chose the preferred Cookie Management.
Benvenuti nel sito Gibitre Instruments
Per continuare scegliete la Gestione dei Cookie che preferite.
Cookies policy
For detailed informations about the use of cookies, please read our
Cookies Management Practice.

Gestione dei Cookies
Per maggiori informazioni sui cookie utilizzati consultate la nostra
Procedura sull'Uso dei Cookies.